High-pitch CT pulmonary angiography (CTPA) with ultra-low contrast medium volume for the detection of pulmonary embolism: a comparison with standard CTPA

Eur Radiol. 2024 Mar;34(3):1921-1931. doi: 10.1007/s00330-023-10101-8. Epub 2023 Sep 1.

Abstract

Objective: To investigate the feasibility and image quality of high-pitch CT pulmonary angiography (CTPA) with reduced iodine volume in normal weight patients.

Methods: In total, 81 normal weight patients undergoing CTPA for suspected pulmonary arterial embolism were retrospectively included: 41 in high-pitch mode with 20 mL of contrast medium (CM); and 40 with normal pitch and 50 mL of CM. Subjective image quality was assessed and rated on a 3-point scale. For objective image quality, attenuation and noise values were measured in all pulmonary arteries from the trunk to segmental level. Contrast-to-noise ratio (CNR) was calculated. Radiation dose estimations were recorded.

Results: There were no statistically significant differences in patient and scan demographics between high-pitch and standard CTPA. Subjective image quality was rated good to excellent in over 90% of all exams with no significant group differences (p = 0.32). Median contrast opacification was lower in high-pitch CTPA (283.18 [216.06-368.67] HU, 386.81 [320.57-526.12] HU; p = 0.0001). CNR reached a minimum of eight in all segmented arteries, but was lower in high-pitch CTPA (8.79 [5.82-12.42], 11.01 [9.19-17.90]; p = 0.005). Median effective dose of high-pitch CTPA was lower (1.04 [0.72-1.27] mSv/mGy·cm; 1.49 [1.07-2.05] mSv/mGy·cm; p < 0.0001).

Conclusion: High-pitch CTPA using ultra-low contrast volume (20 mL) rendered diagnostic images for the detection of pulmonary arterial embolism in most instances. Compared to standard CTPA, the high-pitch CTPA exams with drastically reduced contrast medium volume had also concomitantly reduced radiation exposure. However, objective image quality of high-pitch CTPA was worse, though likely still within acceptable limits for confident diagnosis.

Clinical relevance: This study provides valuable insights on the performance of a high-pitch dual-source CTPA protocol, offering potential benefits in reducing contrast medium and radiation dose while maintaining sufficient image quality for accurate diagnosis in patients suspected of pulmonary embolism.

Key points: • High-pitch CT pulmonary angiography (CTPA) with ultra-low volume of contrast medium and reduced radiation dose renders diagnostic examinations with comparable subjective image quality to standard CTPA in most patients. • Objective image quality of high-pitch CTPA is reduced compared to standard CTPA, but contrast opacification and contrast-to-noise ratio remain above diagnostic thresholds. • Challenges of high-pitch CTPA may potentially be encountered in patients with severe heart failure or when performing a Valsalva maneuver during the examination.

Keywords: Computed tomography pulmonary angiography; Dual-source pulmonary angiography; High-pitch mode; Motion artifacts; Pulmonary embolism.

MeSH terms

  • Angiography / methods
  • Computed Tomography Angiography / methods
  • Contrast Media
  • Humans
  • Hypertension, Pulmonary*
  • Pulmonary Artery / diagnostic imaging
  • Pulmonary Embolism* / diagnostic imaging
  • Radiation Dosage
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods

Substances

  • Contrast Media